What is DICOM?
DICOM is the standard that lets scanners, archives and viewers from different makers exchange medical images. If you were handed a CD or a download from a hospital and found files ending in .dcm, this is the format they use.
Last reviewed 18 October 2026. Written by the dicompeek editors; sources are listed at the end.
The short answer
DICOM stands for Digital Imaging and Communications in Medicine. It is both a file format and a set of rules for sending images over a network. It is developed by the DICOM Standards Committee, with NEMA (the National Electrical Manufacturers Association) serving as its secretariat, and it is also published by ISO as ISO 12052. Its predecessor was the ACR/NEMA standard “Digital Imaging and Communications”, from 1988. The full text is free to read at dicom.nema.org. “DICOM” is a registered trademark of NEMA.
CT, MR, X-ray, ultrasound, nuclear medicine and many other devices write DICOM. So do the systems that store and show their output (the archives hospitals call PACS), which is why one viewer can usually open images from a scanner of any brand.
What is inside a .dcm file
A DICOM file on disk, called a “Part 10 file” after the part of the standard that defines it, is built like this:
- A 128-byte preamble, normally empty, followed by the four letters
DICM. A file without them is probably not a Part 10 file at all. - File meta information: a few elements that say what kind of object follows and, importantly, how it is encoded (the transfer syntax; see transfer syntaxes explained).
- The data set: a list of data elements. Each has a tag, a value representation (VR, the data type), a length and a value.
A tag is a pair of numbers written like (0010,0010), which is Patient's Name. (0008,0060) is Modality,(0028,0010) is Rows, and (7FE0,0010) is Pixel Data: the picture itself is just one element among many. Tags in odd-numbered groups are private: a manufacturer can put anything there. Elements can also hold sequences of items, which is how DICOM stores nested information such as a list of referenced images. The DICOM tag viewer shows all of them for a file.
More than a picture
A DICOM image keeps the information needed to use the pixels properly, not only to look at them:
- Real measurements. A CT slice stores raw values plus a rescale slope and intercept that turn them into Hounsfield units. See window and level basics.
- Geometry. Pixel spacing, slice thickness and position tell software how large a pixel is and where the slice sits in the body, so measurements and reformatting are possible.
- Bit depth. 12 to 16 bits per pixel is normal for CT and MR, far more than a JPEG's 8. Converting to JPG discards the rest; see DICOM to JPG for what changes.
- Identity and context. Patient, study, dates, device, protocol. This is also why DICOM files must be treated as patient data.
Patient, study, series, image
The standard models the data as a hierarchy. A patient has studies (one visit or order, for example “CT head”). A study has series (one acquisition or reconstruction, for example “axial, 5 mm”). A series has instances, the individual images or other objects. Each study, series and instance carries a globally unique identifier (a UID, a string of numbers separated by dots). A CT scan of 300 slices is therefore usually 300 files with the same Study and Series UID and different SOP Instance UIDs and Instance Numbers. A viewer groups files by Series Instance UID and sorts them by Instance Number, which is how the dicompeek viewer builds a stack.
Why the files look odd
- Names such as
IM000001orI1000000, or no extension at all, are common, especially on discs. The content is what matters, not the name. See how to open a .dcm file. - A file named
DICOMDIRis an index, not an image: what is a DICOMDIR. - Some files cannot be shown by every program because the pixel data is compressed in a format the program lacks.
The parts of the standard
The standard is published as numbered parts. Four are useful to know when you meet a file:
| Part | Covers |
|---|---|
| PS3.3 | Information object definitions: what a CT image, an MR image or a report must contain |
| PS3.5 | Data structures and encoding, including transfer syntaxes |
| PS3.6 | The data dictionary: every standard tag, its name and VR |
| PS3.10 | Media storage and the file format |
Other parts cover the network protocol, web services (DICOMweb), conformance, security profiles and more.
Handling DICOM files safely
Treat every file as containing a patient's name, ID and birth date, because most do. Only open or share files you have the right to. The dicompeek tools read files in your browser and do not upload them. To remove identifying data before sharing, seeanonymize DICOM, and read its limits first.
Sources
- DICOM PS3.1, Introduction and Overview
- DICOM PS3.10 chapter 7, DICOM File Format (preamble, DICM prefix, file meta information)
- DICOM PS3.5 chapter 7, The Data Set (data elements: tag, VR, length, value)
- DICOM PS3.6 chapter 6, Registry of DICOM Data Elements
- dicomstandard.org, About DICOM (NEMA as secretariat, ISO 12052)
- DICOM PS3.1 chapter 2, Normative References (ACR/NEMA 300, 1988)
Not for clinical diagnosis. See the medical disclaimer.